Myron L 9PTK with FCE Free Chlorine User Manual

Page 30

background image

26

c.

Press

twice to skip calibration adjustment and accept

the new tempco (3 times if in TDS mode). You are now ready to

measure samples with your new temperature compensation factor.

2. Disabling Temperature Compensation

a.

Select User mode (ref. Procedure to Select a Solution, pg. 25).

b.

With “User” selected, press

. If the display does not

show .00%/°C, hold

long enough to bring the tempco to

.00%/°C (see Figure 3).

c.

Press

twice (3 times if in TDS mode). Temperature

compensation is now disabled (=0) for measurements in User

mode.

3. User Programmable Conductivity to TDS Ratio

This feature allows you to select a custom conductivity to TDS conversion

ratio within the range of 0.20-7.99 for User mode measurements.

To determine the conversion ratio for a custom solution of known

TDS ppm value, measure the solution conductivity at 25ºc with the

Ultrameter III and divide the ppm value by the µS value. For example,

a solution of known 75 ppm TDS and measured 100 µS conductivity at

25ºC would have a conversion ratio of 75/100 or 0.75. Enter the new

conversion ratio as follows: