Myron L 9PTK with FCE Free Chlorine User Manual
Page 51

47
Follow the steps below to set either the KCl, NaCl or 442 calibration
factor to the User solution mode.
1.
Press measurement key desired to be “Linked”, i.e.,
,
or
.
2.
Place the Ultrameter III in User mode (ref. SOLUTION
SELECTION, pg. 24, for selecting the User mode).
3.
Press
arrow key until the menu “linc” appears (see
Figure 27).
4.
Press
key. The instrument will display “sel” and the
“User” Icon (see Figure 28).
Any additional display of KCl, NaCl or 442 icons indicates
a “Linc” between the User solution and the other solution
displayed.
5.
Press
or
keys to select a Standard Solution to be
linked to the User mode calibration constant. In Figure 29, pg.
48, the display indicates that “User” is linked to “Kcl”.