I. Measuring ORP

The Ultrameter III features the ability to measure the activity of

oxidizing or reducing chemicals in solution as ORP mV. The instrument

also includes an innovative Free Chlorine Equivalent (FC

E

) feature

(Measuring Free Chlorine Using FC

E

, pg. 22) that uses ORP and pH to

measure free available chlorine (FAC) concentration in ppm. ORP mV

and ppm of free available chlorine (FAC) are the two most commonly

used sanitizer units of measure in water quality management.

1. ORP / FC

E

Mode Selection

The Ultrameter III allows the user to choose between measuring

oxidizing sanitizers using either ORP mV or as parts per million (ppm)

of equivalent free chlorine. Use ORP to directly measure the oxidizing

power of all sanitizers like ozone, bromine, peracetic acid or chlorine. Use

FC

E

to measure the strength of oxidizing sanitizers as ppm of equivalent

free chlorine. To select between ORP and Free Chlorine modes:
1.

Press

.

2.

Press and hold

for approximately 3 seconds.

The current preference for ORP units of measure is displayed. Factory

setting for this preference is ORP mV.

3.

Press the

or

keys to toggle between mV (standard

ORP mode) and FC

E

ppm. The setting chosen is displayed.

PPM

4.

Press any parameter key to exit ORP unit preference selection

or let the unit time out. ORP unit preference will be saved.