beautypg.com

Myron L 9PTK with FCE Free Chlorine User Manual

Page 39

background image

35

solution before adjusting the gain with acid or base buffers, i.e., 4 and/

or 10, etc.

a. pH Zero Calibration

1.

Rinse sensor well and cell cup 3 times with 7 buffer solution.

2.

Refill both sensor well and cell cup with 7 buffer solution.

3.

Press

to verify the pH calibration. If the display shows

7.00, skip the pH Zero Calibration and proceed to section b.

pH gain Calibration.

4.

Press

to enter calibration mode. The “cal”, “bUFFer

and “7” annunciators will appear (see Figure 6). Displayed

value will be the uncalibrated sensor.

nOte: If a wrong buffer is added (outside of 6-8 pH), “7”and “bUFFer

will flash, and the Ultrameter III will not adjust. The uncalibrated pH

value displayed in step 4 will assist in determining the accuracy of the

pH sensor. If the pH reading is above 8 with pH 7 buffer solution, the

sensor well needs additional rinsing or the pH sensor is defective and

needs to be replaced.

5.

Press

or

until the display reads “7.00”.

nOte: Attempted calibration of >1 pH point from factory calibration will

cause “Fac” to appear. This indicates the need for sensor replacement

(ref. Troubleshooting pg. 54) or fresh buffer solution. The “FAC” internal

electronic calibration is not intended to replace calibration with pH

buffers. It assumes an ideal pH sensor Each “FAC” indicates a factory

setting for that calibration step (i.e., 7, acid, base).

You may press

to accept the preset factory value, or you may

reduce your variation from factory setting by pressing

or

.

6.

Press

to accept the new value. The pH Zero Calibration