Myron L 9PTK with FCE Free Chlorine User Manual
Page 38

34
17.
Using the pipette, add a second 100 µL of Reagent H1 to the
cell, close the cell extender cap securely to prevent any solution
from spilling out, then agitate and hold. Repeat as prompted.
“P
r
es cal” will display.
18.
Press cal to advance to first titration prompt. “add” alternating
with either “edta lc” (Reagent H2) or “edta hc” (Reagent
H3) will display; then “agit t1” alternating with “hOld” will
display. The number after “t” indicates which titration point you
are on. The number “1” here indicates this is the first titration
point. Subsequent titration points are sequenced numerically: 2,
3, 4, etc.
19.
Tap the cell extender cap to dispel any solution clinging to the
cap back into the cell. Carefully open the cell extender cap so
as not to spill any solution.
20.
Change the pipette tip.
21.
Using the pipette, add 100µL of Reagent H2 if “edta lc” is
displayed, or H3 if “edta hc” is displayed.
22.
Close the cell extender cap securely to prevent any solution
from spilling out. Then agitate and hold. Repeat as prompted.
“P
r
es cal” will display.
23.
Press cal to advance to the next titration point. Continue adding
100µL of Reagent H2 or H3 using the pipette, agitating and
holding the unit as prompted, then pressing cal as prompted.
A hardness value will display when the titration measurement
is complete.
24.
Press cal. The “cal” icon will display.
25.
Adjust the displayed value to “200” using UP or DOWN key. The
unit will not allow you to scroll up or down to “200” if it is beyond
±10% of the reading. Ref. Total Return to Factory Settings “Fac
sel”, pg. 43 to reset the Hardness Calibration to the “Fac”
setting along with all other calibrations.
26.
Press cal to accept.
The unit will automatically power off after the period of inactivity
defined in the AUTO OFF setting (ref. AUTO OFF, pg. 45).
7. pH Calibration
imPOrtant: Always “zero” your Ultrameter III with a pH 7 buffer