Myron L 9PTK with FCE Free Chlorine User Manual
Page 32
28
CALIBRATION INTERVALS TABLE:
Function
Kcl, nacl or 442
User
Cond
gain Only
Tempco, then gain
Res
Done in Conductivity
Done in Conductivity
TDS
gain Only
Tempco, Ratio, then gain
pH
7, acid and/or base
ORP
Zero set with pH 7
automatically
Alkalinity
gain Only
Hardness
gain Only
a. Starting Calibration
For Alkalinity and Hardness, calibration is begun by pressing
when the titration is complete and a value displays.
For Conductivity, TDS or pH, calibration is begun by pressing
while measuring Conductivity, TDS or pH. Measuring continues, but
the “cal” icon is on, indicating calibration is now changeable. The
reading is changed with the
and
keys to match the
known value.
The calibration for each of the 4 solution types may be performed in
either conductivity or TDS mode.
Once in CAL mode, the
key becomes an “ACCEPT” key for the
new calibration.
For pH, at each point, pressing
accepts the new calibration
value and steps you to the next adjustment.
To bypass a calibration, simply press
to accept the present
value as-is.
b. Leaving Calibration
Pressing
accepts calibration and exits CAL mode if there are no
more steps. Calibration is complete when the “cal” icon goes out.
Pressing any measurement key during calibration cancels changes
not yet accepted and exits calibration mode. Leaving pH after the 2nd
buffer results in the same gain being entered in place of the 3rd buffer.
2. Calibration Limits
There are calibration limits for all functions for which calibration is
required. For Alkalinity and Hardness, the 9P will not allow you to