Campbell Scientific OBS-3A Turbidity and Temperature Monitoring System User Manual

Page 59

background image

OBS-3A Turbidity and Temperature Monitoring System

b. If the OBS-3A still fails to respond, try changing baud rate speeds by

selecting Attempt to find the baud rate to communicate. If this fails,
switch the PC back to 115.2 kB and go to the next step.

c. Reconnect the cable and try to connect again.

d. Replace the main batteries; see Section 6.1.2, Battery Installation, and

try to connect again.

e. If you have a survey cable, connect instrument to external power and

try to connect again.

f. Remove the unit from the pressure housing and press and release the

RESET button. Try to connect again.

51