1. effects of disaggregation, 2. sample durations for sediment calibrations – Campbell Scientific OBS-3A Turbidity and Temperature Monitoring System User Manual

Page 56

background image

OBS-3A Turbidity and Temperature Monitoring System

10

20

30

40

50

Sediment Concentration (mg/l)

0

10

20

30

40

50

60

T

u

rb

id

ity

(NT

U

)

Hand Shaking

15-min. Sonic Bath

3-min. Sonic Probe

FIGURE 7-1. Effects of disaggregation

Preparation

1. Clean containers and glassware with detergent and rinse with filtered

water.

2. Perform the calibration under fluorescent lighting.

3. Based on the material, select the appropriate sample duration from TABLE

7-2.

TABLE 7-2. Sample Durations for

Sediment Calibrations

Sediment Seconds

Clay 10

Silt 20

Fine Sand

40


4. Open the calibration the screen under OBS-3A Configuration |

Operation | Calibration | Calibrate and select mg/L.

5. Enter sediment concentration values.

6. After each addition of sediment, compute mg/L or ppm with the equations

given below.

48