OBS-3A Turbidity and Temperature Monitoring System

5.1 Measurement Range

Turbidity (AMCO Clear):

0.4 to 4,000 NTU

1

Mud (D

50

=20 µm):

0.4 to 5,000 mg/L

Sand (D

50

=250 µm):

2 to 100,000 mg/L

Pressure

2

:

0 to 10, 20, 50, 100, or 200 m

Temperature:

to

35°C

Conductivity (salinity):

0 to 65 mS/cm (40 PSU, o/oo)

1

0 to 100, 0 to 250, 0 to 500, 0 to 1000, 0 to 2000, and 0 to 4000 NTU
ranges are available.

2

Range depends on pressure sensor option chosen.

5.2 Accuracy

Turbidity (AMCO Clear,
0 to 2,000 NTU):
<2.0%

Mud (0.4 to 4,000 mg/L): 2.0%

of

reading

Sand (0.4 to 60,000 mg/L): 3.5%

of

reading

Pressure:

±0.5% full scale

Temperature: ±0.5°C

Conductivity: 1%

5.3 OBS Sensor

Frequency: 5

Hz

Drift over time: <2%

per

year

Drift over temperature:

0.05% per °C

5.4 Other Data

Maximum size sample: 2048

Sampling rate:

1 to 25 Hz

Maximum data rate: 25

Hz

Data capacity: 8

Mbytes

Maximum number
of data lines:
200,000

5