1 procedure for removing rh chip, 5 cr1000m module, Procedure for removing rh chip – Campbell Scientific ET107 Evapotranspiration Monitoring Station User Manual

Page 76: Cr1000m module

background image

ET107 Weather Station

2 years

Replace 034B-ET’s vane potentiometer if needed (call for part number and

price). To send the sensor to Campbell Scientific for potentiometer

replacement call for Return Material Authorization number or contact

Campbell Scientific for potentiometer cost to replace potentiometer in the

field.

Replace enclosure gasket if necessary. Enclosure can be sent in to

Campbell Scientific for gasket replacement – call for Return Material

Authorization number or contact Campbell Scientific for gasket cost to

replace gasket in the field.

3 years

Send the solar radiation sensor (pyranometer) to Campbell Scientific for

calibration. Call Campbell Scientific for a Return Material Authorization

number. Sensor cannot be calibrated in the field. (Some users recommend

calibrating this sensor on a yearly basis.)

4-5 years

Check sensor cables (as well as all other cables) for cracking,

deterioration, proper routing, and strain relief. Replace as required.

Check enclosure relative humidity sensor (pn 10070). To check this

sensor, take the lid off the enclosure during routine desiccant replacement

and leave it off for 5 to 10 minutes before putting in new desiccant. While

the lid is off the enclosure, compare the internal enclosure humidity to the

air humidity. Replace if > 10% off.

4.1.4.1 Procedure for Removing RH Chip

1. Unscrew the protective cap (see FIGURE 4-1).

2. Hold the plastic sides of the RH chip and unplug it.

To prevent scratching, avoid touching the silver chip, and

handle the RH chip with care.

3. Rinse off the RH chip with distilled water or dispose of the old RH chip.

4. Hold the sides of the rinsed or new chip and plug it in.

5. Screw on the protective cap.

4.1.5 CR1000M Module

The CR1000M Module contains a lithium battery that operates the clock and

SRAM when the module is not powered. The CR1000M does not draw power

from the lithium battery while it is powered by a 12 Vdc supply. In an ET107

stored at room temperature, the lithium battery should last approximately 10

years (less at temperature extremes). Where the ET107 is powered most or all

of the time, the lithium cell should last much longer.

CAUTION