1. exploded view of hmp60-ets (as shipped) – Campbell Scientific ET107 Evapotranspiration Monitoring Station User Manual

Page 75

background image

ET107 Weather Station

6 months

Replace 034B-ET’s anemometer bearings and reed switch if operating

under harsh conditions, such as constant high winds, blowing dust, and/or

salt spray contamination.

1 year

Replace 034B-ET wind speed (anemometer) bearings (pn 3648). To send

the sensor to Campbell Scientific for bearing replacement call for Return

Material Authorization number or contact Campbell Scientific for bearing

cost to replace bearings in the field.

Replace 034B-ET wind speed (anemometer) reed switch (pn 13764) if

needed. To send the sensor to Campbell Scientific for reed switch

replacement call for Return Material Authorization number or contact

Campbell Scientific for reed switch cost to replace reed switch in the field.

Check calibration of the HMP60-ETS Temp/RH probe. Sensor will tend to

drift up over time giving readings higher than 100%. Replace RH chip

(pn 9598) if necessary (see FIGURE 4-1 and Section 4.1.4.1, Procedure

for Removing RH Chip).

ET107 stations began shipping January 2014 with an updated gill

radiation shield with knurled thumb screws, using the HMP60-

ETS Temp/RH sensor. All information in this document is also

pertinent for the HMP50-ET, HMP60-ET, and HMP60-ETR

sensors.

Replace desiccant in enclosure housing as needed.

FIGURE 4-1. Exploded view of the temperature and RH sensor

(adapter not shown)

NOTE

Protective Cap

and Filter

Shipping Cap

(remove prior

to installation)

9598 RH Chip

65