Appendix B. Changing the HC2S3 Settings

B.2 Software and Hardware Requirements

For temperature (Analog Output 2), the HC2S3 default range is -40 to +60°C
for 0 to 1V. Changing the range requires Rotronic HW4 Software (Version
2.1.0 or higher), and the Rotronic model AC3001 USB adapter cable. Power to
the probe is provided by the USB port.

Prior to using the AC3001 cable, the ROTRONIC USB driver
must be installed on the PC. Both the driver and the installation
instructions (document E-M-HW4v3-Main) are located on the
HW4 CD.

IMPORTANT

B.3 Changing the Temperature Range

Install the HW4 software and drivers for the AC3001 USB cable on the PC.
Connect the HC2S3 probe to the AC3001 cable, making sure the connectors
are properly aligned before tightening the knurled ring. Plug the AC3001 cable
into a USB port on the computer.

From the main screen, click on the “devices and groups”, search for “master
devices”, USB masters.

HW4 should find the probe, and show the current values:

B-2