5 measuring probes with long cables, Measuring probes with long cables – Campbell Scientific HC2S3 Temperature and Relative Humidity Probe User Manual

Page 25

background image

Model HC2S3 Temperature and Relative Humidity Probe

2: Volt (Diff) (P2)
1: 1

Reps

2: 25

2500 mV 60 Hz Rejection Range

3: 2

DIFF Channel

4: 4

Loc [ RH ]

5: 0.1

Multiplier

6: 0

Offset


3: If (X<=>F) (P89)
1: 4

X Loc [ RH ]

2: 3

>=

3: 100

F

4: 30

Then Do


4: If (X<=>F) (P89)
1: 4

X Loc [ RH ]

2: 4

<

3: 103

F

4: 30

Then Do


5: Z=F x 10^n (P30)
1: 100

F

2: 0

n, Exponent of 10

3: 4

Z Loc [ RH ]


6: End (P95)

7: End (P95)

8: If time is (P92)
1: 0

Minutes (Seconds --) into a

2: 60

Interval (same units as above)

3: 10

Set Output Flag High (Flag 0)


9: Set Active Storage Area (P80)
1: 1

Final Storage Area 1

2: 101

Array ID


10: Real Time (P77)
1: 1220 Year,Day,Hour/Minute (midnight = 2400)

11: Average (P71)
1: 1 Reps
2: 3 Loc [ AirTC ]

12: Sample (P70)
1: 1 Reps
2: 4 Loc [ RH ]

7.5 Measuring Probes with Long Cables

For cable lengths longer than 6.1 m (20 ft), Campbell Scientific recommends
measuring the voltage signals using differential inputs as discussed below.
Connections for differential inputs are given in TABLE 7-2.

19