4 programming, Programming, 2. connections – Campbell Scientific HC2S3 Temperature and Relative Humidity Probe User Manual
Page 19: Differential measurements, Table 7-2

Model HC2S3 Temperature and Relative Humidity Probe
TABLE 7-2. Connections for Differential Measurements
Color
Description
CR1000,
CR3000,
CR800,
CR5000,
CR23X
CR10X
CR10,
CR510,
CR500
Brown
Temperature
Signal
Differential
Input – H
Differential
Input – H
Jumper to
Yellow
Temperature Signal
Reference
Differential
Input – L
Differential
Input – L
White
Relative Humidity Signal
Differential
Input – H
Differential
Input – H
Yellow Signal
Reference
Differential
Input – L
Differential
Input – L
Gray Power
Ground
G G
Clear Shield
G
Green Power
12V/*SW12V
12V/*SW12V
*CR10X Power Control if
using SW 12V
Jumper from
SW 12V
CTRL to
Control Port
7.4 Programming
This section is for users who write their own datalogger
programs. You do not need to read this section if using our Short
Cut Program Generator, or connecting the probe to a prewired
enclosure, RAWS-P station, RAWS-F station, or CWS900
Wireless Sensor Interface. Our prewired enclosures, RAWS-P
stations, and RAWS-F stations include a datalogger program.
Refer to the Wireless Sensor Manual for programming
information if using the probe with a CWS900.
NOTE
The temperature and relative humidity signals from the HC2S3 are measured
using either single-ended voltage instructions (VoltSE() in CRBasic or
Instruction 1 in Edlog) or differential voltage instructions (VoltDiff() in
CRBasic or Instruction 2 in Edlog). Differential measurements are
recommended for cables longer than 6.0 m (20 ft) as discussed in Section 7.5.
When the probe is connected to a CS110 Electric Field Meter,
the probe is measured by the CS110's internal CR1000
datalogger module using VoltSE() instructions. Relative
humidity and temperature signals are measured on single-ended
channels 1 and 2, respectively. The 250 µs integration should be
used in the VoltSE() instructions.
NOTE
13