Appendix B. Changing the HC2S3
Settings

B.1 HC2S3 Default Settings

The HC2S3 probe has the following default settings, which can be changed as
described in the following sections. Additional information can be found in
Rotronic’s User Manual: E-M-HC2 Probes-VXXXX, which can be
downloaded from Rotronic's website

www.rotronic-usa.com

.

Default Settings:

Configurable Settings

Factory Default

Unit system (Metric or English)

Metric

Psychrometric calculation

None

Output 1 parameter, scale and unit

Humidity: 0..100% RH

Output 2 parameter, scale and unit

Temperature: -40...+60 deg C

Communications Protocol

RO-ASCII

RS-485 Address

0

Device name

Probe type

Humidity / temperature adjustment
Device write protection

Disabled

Limit humidity output to 100% RH Disabled
Out-of-limit value digital alarm

Disabled

Data recording

Enabled (loop mode - 10 min interval)

Automatic humidity sensor test

Disabled

Humidity sensor drift compensation Disabled
Fail safe mode

Disabled

Simulator mode

Disabled

Digital Interface:

Interface Type: UART (Universal Asynchronous Receiver Transmitter)

Organization: Dialog, duplex

Default Configuration:

Baud rate: 19200

Parity:

none

Data

bits:

8

Stop fits: 1

Flow Control: none

Logical Levels:

Logical 0: <= 0.3V*VDD

Logical 1: <= 0.8V*VDD

B-1