Model HC2S3 Temperature and Relative Humidity Probe

The HC2S3 output scale is 0 to 1000 mV for the temperature range of
-40

° to +60°C and 0 to 1000 mV for the relative humidity range of 0 to 100%.

Multipliers and Offsets for the measurement instructions to convert the
measurement result (mV) to temperature and relative humidity are shown in
TABLE 7-3 and TABLE 7-4.

TABLE 7-3. Temperature

Units

Multiplier

(degrees mV

-1

)

Offset

(degrees)

Celsius 0.1

-40

Fahrenheit 0.18

-40

TABLE 7-4. Humidity

Units

Multiplier

(% mV

-1

)

Offset

(%)

Percent 0.1

0

Fraction 0.001

0

7.4.1 Example Programs using Single-Ended Measurement Instructions

The example programs for the CR1000 and CR10X use the SW12V terminal to
switch power to the probe, delay for 3 seconds and measure the output voltages
using single-ended measurement instructions.

Relative humidity and temperature (deg C) are measured on single-ended input
channels 1 and 2 respectively. The program sets relative humidity equal to 100
if the measured value is greater than 100 but less than 103%. Values greater
than 103% are not set equal to 100, and indicate a problem with the sensor or
its calibration.

TABLE 7-5. Wiring for Single-ended

Measurement Examples

Color Description CR1000 CR10(X)

Brown

Temperature

SE 2

SE 2

White

Relative Humidity

SE 1

SE 1

Yellow Signal

Reference

AG

Gray Power

Ground

AG

Clear Shield

G

Green Power SW12V

SW12V

Jumper from

SW 12V CTRL

to Control Port

14