6.2 Relative Humidity Sensor

Sensor: ROTRONIC

Hygromer

®

IN1

Relative Humidity
Measurement Range:

0 to 100% non-condensing

RH Output Signal Range:

0 to 1 Vdc

Accuracy at 23

°C:

±0.8% RH with standard configuration

settings

Typical Long Term Stability:

<1% RH per year

Sensor Time Constant (63% of a
35 to 80% RH step change
(1 m/s air flow at sensor)):

≤22 s with PE filter, ≤30 s with Teflon
filter

RH Accuracy over Temperature:

The black outer jacket of the cable is Santoprene

®

rubber.

This compound was chosen for its resistance to
temperature extremes, moisture, and UV degradation.
However, this jacket will support combustion in air. It is
rated as slow burning when tested according to U.L. 94
H.B. and will pass FMVSS302. Local fire codes may
preclude its use inside buildings.