beautypg.com

1 temperature sensor, Temperature sensor – Campbell Scientific HC2S3 Temperature and Relative Humidity Probe User Manual

Page 15

background image

Model HC2S3 Temperature and Relative Humidity Probe

Power Consumption:

<4.3 mA @ 5 V

<2.0 mA @ 12 V

Supply Voltage (using CSI cable): 5 to 24 Vdc (12 Vdc recommended)

Start-up time:

1.5 s typical (Rotronic specification,

Campbell Scientific recommends 2 s at

60°C, 3 s at 0°C, 4 s at -40°C)

Maximum Startup Current:

<50 mA during 2 µs

Maximum Lead Length:

300 m (1000 ft) with 12 V power, 3 m

(10 ft) with 5 V power

Analog outputs

Offset at 0 V:

±3 mV (maximum)

Deviation from Digital Signal: < ±1 mV (0.1°C, 0.1% RH)

6.1 Temperature Sensor

Sensor:

PT100 RTD, IEC 751 1/3 Class B, with

calibrated

signal

conditioning

Temperature Measurement Range: -50

° to +100°C (default -40° to + 60°C)

Temperature Output Signal Range: 0 to 1.0 V

Accuracy at 23°C:

±0.1°C with standard configuration

settings

Long Term Stability: <0.1°C/year

Sensor Time Constant (63% step
change (1 m/s air flow at sensor)):

≤22 s with PE filter, ≤30 s with Teflon
filter

Temperature Accuracy over Temperature:

9