beautypg.com

7 100 ohm prt in 3 wire half bridge – Campbell Scientific CR510 Basic Datalogger User Manual

Page 84

background image

SECTION 7. MEASUREMENT PROGRAMMING EXAMPLES

7-6

7.7 100 OHM PRT IN 3 WIRE HALF

BRIDGE

The temperature measurement requirements in
this example are the same as in Section 7.8. In
this case, a three wire half bridge, Instruction 7,
is used to measure the resistance of the PRT.
The diagram of the PRT circuit is shown in Fig.
7.7-1.

As in the example in Section 7.8, the excitation
voltage is calculated to be the maximum
possible, yet allow the +25 mV measurement
range. The 10 kohm resistor has a tolerance of

±

1%; thus, the lowest resistance to expect from

it is 9.9 kohms. We calculate the maximum
excitation voltage (V

x

) to keep the voltage drop

across the PRT less than 25 mV:

0.025V > V

x

115.54/(9900+115.54);
V

x

< 2.17 V

The excitation voltage used is 2.1 V.

The multiplier used in Instruction 7 is
determined in the same manner as in Section
7.8. In this example, the multiplier (R

f

/R

0

) is

assumed to be 100.93.

The 3 wire half bridge compensates for lead
wire resistance by assuming that the resistance
of wire A is the same as the resistance of wire
B. The maximum difference expected in wire

resistance is 2%, but is more likely to be on the
order of 1%. The resistance of R

s

calculated

with Instruction 7, is actually R

s

plus the

difference in resistance of wires A and B. The
average resistance of 22 AWG wire is 16.5
ohms per 1000 feet, which would give each 500
foot lead wire a nominal resistance of 8.3 ohms.
Two percent of 8.3 ohms is 0.17 ohms.
Assuming that the greater resistance is in wire
B, the resistance measured for the PRT (R

0

=

100 ohms) in the ice bath would be 100.17
ohms, and the resistance at 40

°

C would be

115.71. The measured ratio R

s

/R

0

is 1.1551;

the actual ratio is 115.54/100 = 1.1554. The
temperature computed by Instruction 16 from
the measured ratio would be about 0.1

°

C lower

than the actual temperature of the PRT. This
source of error does not exist in the example in
Section 7.8, where a 4 wire half bridge is used
to measure PRT resistance.

The advantages of the 3 wire half bridge are
that it only requires 3 lead wires going to the
sensor and takes 2 single-ended input
channels, whereas the 4 wire half bridge
requires 4 wires and 2 differential channels.

A terminal input module (Model 3WHB10K) can
be used to complete the circuit in Figure 7.7-1.

CR510

FIGURE 7.7-1. 3 Wire Half Bridge Used to Measure 100 ohm PRT