29 inw ps9105e, 105 sdi-12 recorder – Campbell Scientific CR510 Basic Datalogger User Manual
Page 115

SECTION 9. INPUT/OUTPUT INSTRUCTIONS
9-11
PARAM.
DATA
NUMBER
TYPE
DESCRIPTION
01:
2
Repetitions Hit C (--) to
skip repeat of
excitation
02:
2
Single-ended channel
for first measurement
03:
2
Excitation Channel
04:
2
Start frequency of
sweep (100'S of Hz)
05:
2
End frequency of
sweep (100'S of Hz)
06:
4
# Cycles to measure (0
means none)
07:
4
Delay before excitation
applied (0.01 sec units)
08:
4
Input location (1/T
2
), T
in ms
09:
FP
Multiplier
10:
FP
Offset
Input locations altered: 1 per repetition
*** 29 INW PS9105E ***
FUNCTION
The Instrumentation Northwest PS9105
Enhanced Pressure Transducer is used to
measure water level. This instruction excites
the PS9105 with a single excitation channel and
measures the sensor’s output on two
consecutive differential analog input channels.
The pressure and sensor temperature are
written into two consecutive input locations
starting at the location specified in parameter 3.
Each PS9105 has 20 enhanced measurement
parameters that are documented on the
calibration sheet. Enter these parameters into
Instruction 29.
PARAM.
DATA
NUMBER
TYPE
DESCRIPTION
01:
2
Differential channel
number for first
measurement
02:
2
Excitation channel
number
03:
4
Input location number
04:
FP
Pressure Range (psig)
(enh. par. 1)
05:
FP
Enhanced Parameter 2
06:
FP
Enhanced Parameter 3
07:
FP
Enhanced Parameter 4
08:
FP
Enhanced Parameter 5
09:
FP
Enhanced Parameter 6
10:
FP
Enhanced Parameter 7
11:
FP
Enhanced Parameter 8
12:
FP
Enhanced Parameter 9
13:
FP
Enhanced Parameter 10
14:
FP
Enhanced Parameter 11
15:
FP
Enhanced Parameter 12
16:
FP
Enhanced Parameter 13
17:
FP
Enhanced Parameter 14
18:
FP
Enhanced Parameter 15
19:
FP
Enhanced Parameter 16
20:
FP
Enhanced Parameter 17
21:
FP
Enhanced Parameter 18
22:
FP
Enhanced Parameter 19
23:
FP
Enhanced Parameter 20
Input location altered: 2
*** 105 SDI-12 RECORDER ***
NOTE: Version 1.2 of the SDI-12
specification has been implemented in the
CR10X. Features added in this version are
marked with a dagger (†) and may not be
supported by older SDI-12 sensors.
Instruction 105 allows data to be collected from
SDI-12 sensors. The sensor’s SDI-12 data line
is connected to a control port (C1 or C2). The
SDI-12 ground should be connected to a "G"
terminal on the CR10X wiring panel. SDI-12
power may be connected to 12V.
If multiple SDI-12 sensors are used, up to ten
sensors may be connected to a single control
port, but each sensor must have a unique
address and requires a separate Instruction
105.
PARAM.
DATA
NUMBER
TYPE
DESCRIPTION
01:
4
Address (0-9, †10-
126)
02:
2
Command (Table 9-8)
03:
2
Control Port (C1-C8)
04:
4
Input Loc.
05:
FP
Mult
06:
FP
Offset
Input locations altered: 1-9 (†1-99), depending
on the SDI-12 sensor
Intermediate locations required: 21