Campbell Scientific CR510 Basic Datalogger User Manual
Page 69

SECTION 5. TELECOMMUNICATIONS
5-3
TABLE 5.1-1. Telecommunications Commands
Command
Description
[F.S. Area]A
SELECT AREA/STATUS - If 1 or 2 does not precede the A to select the
Final Storage Area, the CR510 will default to Area 1. All subsequent
commands other than A will address the area selected. Datalogger
returns Reference, the DSP location; the number of filled Final Storage
locations; Version of datalogger; Final Storage Area; Location of MPTR
(the location number may be 1 to 7 characters long); Errors #1, #2, and
#3 where #1 is the number of E08's, #2 is the number of overrun errors,
and #3 is the number of times the program stopped due to low voltage
(all are cleared by entering 8888A; #2 is also cleared at time of program
compilation); size of total Memory in CR510; the lithium Battery voltage;
and Checksum. All in the following format:
R+xxxxx F+xxxxx Vxx Axx L+xxxxxxx Exx xx xx Mxxxx B+xxxxx Cxxxx
If data is stored while in telecommunications, the A command must be
issued to update the Reference to the new DSP.
[no. of arrays]B
BACK-UP - MPTR is backed-up the specified number of Output Arrays
(no number defaults to 1) and advanced to the nearest start of array.
CR510 sends the Area, MPTR Location, and Checksum:
Ax Lxxxxxxx Cxxxx
[YR:DAY:HR:MM:SS]C
RESET/SEND TIME - If time is entered the time is reset. If only 2
colons are in the time string, HR:MM:SS is assumed; 3 colons means
DAY:HR:MM:SS. If only the C is entered, time is unaltered. CR510
returns year, Julian day, hr:min:sec, and Checksum:
Y:xx Dxxxx Txx:xx:xx Cxxxx
[no. of arrays]D
ASCII DUMP - If necessary, the MPTR is advanced to start of scan.
CR510 sends the number of arrays specified (no number defaults to 1)
or the number of arrays between MPTR and Reference, whichever is
smaller, CRLF, Location, Checksum.
E
End call. Datalogger sends CRLF only.
[no. of loc.]F
BINARY DUMP - Used by CSI software for data retrieval. See
Appendix C.
[F.S. loc. no.]G
MOVE MPTR - MPTR is moved to specified Final Storage location.
The location number must be entered. CR510 sends Area, Location,
and Checksum: Ax Lxxxxxxx Cxxxx
7H or 2718H
REMOTE KEYBOARD - CR510 sends the prompt ">" and is ready to
execute standard keyboard commands (Section OV3).
[loc. no.]I
Display/change value at Input Storage location. CR510 sends the value
stored at the location. A new value and CR may then be sent. CR510
sends checksum. If no new value is sent (CR only), the location value
will remain the same.