beautypg.com

5 tipping bucket rain gage with long leads, 6 100 ohm prt in 4 wire half bridge – Campbell Scientific CR510 Basic Datalogger User Manual

Page 82

background image

SECTION 7. MEASUREMENT PROGRAMMING EXAMPLES

7-4

CR510

FIGURE 7.5-1. Wiring Diagram for Rain Gage with Long Leads

7.5 TIPPING BUCKET RAIN GAGE WITH

LONG LEADS

A tipping bucket rain gage is measured with the
Pulse Count Instruction configured for Switch
Closure. Counts from long intervals will be
used, as the final output desired is total rainfall
(obtained with Instruction 72, Totalize). If
counts from long intervals were discarded, less
rainfall would be recorded than was actually
measured by the gage (assuming there were
counts in the long intervals). Output is desired
in millimeters of precipitation. The gage is
calibrated for a 0.01 inch tip, therefore, a
multiplier of 0.254 is used.

In a long cable there is appreciable capacitance
between the lines. The capacitance is
discharged across the switch when it closes. In
addition to shortening switch life, a transient
may be induced in other wires packaged with
the rain gage leads each time the switch closes.
The 100 ohm resistor protects the switch from
arcing and the associated transient from
occurring, and should be included any time
leads longer than 100 feet are used with a
switch closure.

NOTE: The TE525 and TE525MM
raingages from CSI always have this
resistor installed.

PROGRAM

01:

Pulse (P3)
1:

1

Reps

2:

1

Pulse Input Channel

3:

2

Switch Closure, All Counts

4: 11

Loc [ Precip_mm ]

5:

.254

Mult

6:

0

Offset

7.6 100 OHM PRT IN 4 WIRE HALF

BRIDGE

Instruction 9 is the best choice for accuracy
where the Platinum Resistance Thermometer
(PRT) is separated from other bridge
completion resistors by a lead length having
more than a few thousandths of an ohm
resistance. In this example, it is desired to
measure a temperature in the range of -10 to
40

°

C. The length of the cable from the CR510

to the PRT is 500 feet.

Figure 7.6-1 shows the circuit used to measure
the PRT. The 10 kohm resistor allows the use
of a high excitation voltage and low voltage
ranges on the measurements. This insures that
noise in the excitation does not have an effect
on signal noise. Because the fixed resistor (R

f

)

and the PRT (R

s

) have approximately the same

resistance, the differential measurement of the
voltage drop across the PRT can be made on
the same range as the differential measurement
of the voltage drop across R

f

.

If the voltage drop across the PRT (V

2

) is kept

under 50 mV, self heating of the PRT should be
less than 0.001

°

C in still air. The best

resolution is obtained when the excitation
voltage is large enough to cause the signal
voltage to fill the measurement voltage range.
The resolution of this measurement on the
25mV range is +0.04

°

C. The voltage drop

across the PRT is equal to V

x

multiplied by the

ratio of R

s

to the total resistance, and is

greatest when R

s

is greatest (R

s

=115.54 ohms

at 40

°

C). To find the maximum excitation

voltage that can be used, we assume V

2

equal

to 25 mV and use Ohm's Law to solve for the
resulting current, I.

I = 25 mV/R

s

= 25 mV/115.54 ohms = 0.216 mA