beautypg.com

Ashcroft GC51 - Industrial Rangeable Pressure Transmitter User Manual

Page 8

background image

8

system check and troubleshooting, 4 to 20mA

L. Pressure Connection:

1

4

NPT Female

M. Enclosure:

Material: Aluminum (die cast)
Environmental Rating: IP65 / NEMA 4X

N. Mounting:

Pressure Port Location: Lower (bottom)
Panel mounting: via bracket included
Location: compliant with environmental ratings and where display does
not get directs sunlight

O. Electrical Termination:

Cable Gland: Cable diameter 0.35 to 0.47˝ (9-12 mm)
Terminal block: 14-22 AWG (strand or a solid wire)

1

⁄2

NPT female conduit (optional)

P. Memory Protection:

Permanently stored by EEPROM (nonvolatile memory)

Q. EMC Directive:

CE Complaince:EN61326/1997, A1/1998, A2/2001

R. Operating Temperature: –10 to 60°C (14 to 140°F)

S. Storage Temperature: –20 to 70°C (–4 to 158°F)

T. Vibration: 5g’s 60Hz; x, y, z (2.5 hours for each)

U. Shock: 10g’s 60Hz; x, y, z (3 times forward and backwards each axis)

V. Insulation Resistance: 50Vdc, 100m Ohm minimum

W. Breakdown Voltage: 250Vac for 1 minute

X. Weight: Approx.450g (1.0 lb)

5.

MOUNTING

5.1.

Mounting Location
• Non-hazardous rated area (no combustible gas or liquids)
• Compliant with operating temperature specifications
• Without display receiving direct sunlight
• Compliant with IP65 / NEMA 4X environmental rating