Peripherals and interfaces operational tests, 8 373xxa mm – Anritsu 373XXA User Manual
Page 44
Equipment Required:
Printer Test Fixture (ANRITSU PN: B39553)
Procedure:
Step 1.
Press the
Option Menu
key (Enhancement key group) to dis-
play the OPTIONS menu.
Step 2.
Select DIAGNOSTICS from menu; then in sequence se-
lect: PERIPHERAL TESTS, and then PRINTER IN-
TERFACE. See diagram below.
Step 3.
Follow instructions on the screen until test is completed.
If test fails, proceed to Chapter 5, Troubleshooting.
PERIPHERALS AND INTERFACES
OPERATIONAL TESTS
3-8
373XXA MM
OPTIONS
TRIGGERS
REAR PANEL
OUTPUT
DIAGNOSTICS
MULTIPLE SOURCE
CONTROL
PRESS
SELECT
DIAGNOSTICS
START SELF TEST
READ SERVICE LOG
INSTALLED OPTIONS
PERIPHERAL TESTS
TROUBLESHOOTING
(FOR SERVICE
ONLY)
HW CALIBRATIONS
(FOR SERVICE
ONLY)
PRESS
SELECT
PERIPHERAL TESTS
CRT DISPLAY
FRONT PANEL
EXTERNAL KEY-
BOARD
PRINTER INTERFACE
GPIB INTERFACE
PREVIOUS MENU
PRESS
SELECT