8-3

COVERS

Adjustment and troubleshooting operations require removal of the top
cover. Replacement of some 373XXA assemblies and parts require re-
moval of all covers. The following procedures describe this process.

NOTE

It is only necessary to loosen the 373XXA handle assem-
blies to remove the top, bottom, or side covers. However, if
the front panel is to be removed, remove the handle as-
semblies at this time.

Preliminary:

q Switch the 373XXA power off. Remove the power cord.

Procedure:

Step 1.

Loosen (or remove) the right and left handle assemblies, as

follows:

n Place the 373XXA on its top (bottom-side up).
n Loosen/remove the screws at the sides of the handle as-

semblies.

n If removing handles, pull them away from unit and set

aside.

CAUTION

The green headed screws have Metric
threads.

Step 2.

To remove the top cover:

n Place the 373XXA in normal (top-side up) position.
n Remove the feet from the two top corners at the rear of

the 373XXAB (Figure 7-1).

n Remove the center screw from rear of the top cover.
n Lift and slide the top cover away from the 373XXA.

Step 3.

To remove the bottom cover:

n Place the 373XXA on its top (bottom-side up).
n Remove the feet from the two bottom corners at the rear

of the 373XXA.

n Remove the center screw from rear of the bottom cover.
n Lift and slide the top cover away from the 373XXA.

Step 4.

To remove the left cover:

COVERS

REMOVE AND REPLACE

8-4

373XXA MM