SELF TEST

The self test performs a series of tests that verify that various internal
373XXA circuits are functional and operating properly.

To start the self test:

Step 1.

Press the

Option Menu

key (Enhancement key group) to dis-

play the OPTIONS menu.

Step 2.

Select DIAGNOSTICS from menu; then select START
SELF TEST
(below).

Step 3.

Wait for test sequence to complete. (Once invoked, this test
requires no user interaction or external equipment.)

Upon completion, the net pass/fail result of the self test is shown on
the CRT display. If the 373XXA is in remote-only operation, the re-
sults are reported via the GPIB output buffer. If the self test fails, de-
tailed error messages will be written into the service log.

If self test fails:

q Check the service log to view failure messages.
q Proceed to Chapter 5, Troubleshooting.