3 self test 3-5, Operational tests self test – Anritsu 373XXA User Manual
Page 41

3-3
SELF TEST
The self test performs a series of tests that verify that various internal
373XXA circuits are functional and operating properly.
To start the self test:
Step 1.
Press the
Option Menu
key (Enhancement key group) to dis-
play the OPTIONS menu.
Step 2.
Select DIAGNOSTICS from menu; then select START
SELF TEST (below).
Step 3.
Wait for test sequence to complete. (Once invoked, this test
requires no user interaction or external equipment.)
Upon completion, the net pass/fail result of the self test is shown on
the CRT display. If the 373XXA is in remote-only operation, the re-
sults are reported via the GPIB output buffer. If the self test fails, de-
tailed error messages will be written into the service log.
If self test fails:
q Check the service log to view failure messages.
q Proceed to Chapter 5, Troubleshooting.
OPERATIONAL TESTS
SELF TEST
373XXA MM
3-5
OPTIONS
TRIGGERS
REAR PANEL
OUTPUT
DIAGNOSTICS
MULTIPLE SOURCE
CONTROL
PRESS
SELECT
DIAGNOSTICS
START SELF TEST
READ SERVICE LOG
INSTALLED OPTIONS
PERIPHERAL TESTS
TROUBLESHOOTING
(FOR SERVICE
ONLY)
HW CALIBRATIONS
(FOR SERVICE
ONLY)
PRESS
SELECT