SERVICE LOG ERROR MESSAGES
APPENDIX B
B-6
373XXA MM
Error Message
Description
Corrective Action
STANDARD CAL NOT VALID FOR
WAVEGUIDE
Cannot use the standard method
when calibrating with waveguide.
Use the Offset Short method with
waveguide.
START F FOLLOWS PREVIOUS
STOP F
Start frequency of current band imme-
diately follows stop frequency of previ-
ous band. Cannot be modified.
None.
START MUST BE LESS THAN STOP
Entered start frequency is greater
than the stop frequency.
Re-enter frequency values such that
the start frequency is lower than the
stop frequency.
STEP IS TOO LARGE
Entered harmonic frequency extends
the stop out of range.
Re-enter so that harmonic frequency
is within range.
STOP IS OVER RANGE
Entered value exceeds the instru-
ment’s stop frequency.
Re-enter stop frequency.
SYSTEM NOT CALIBRATED
373XXA is uncalibrated for the se-
lected measurement values.
Perform a measurement calibration.
TOO FEW POINTS, 2 MINIMUM
Entered too few discrete fill points, 2
is minimum.
Re-enter data points.
TOO MANY POINTS, 1601 MAXI-
MUM
Entered too many discrete fill points,
1601 points are the maximum al-
lowed.
Re-enter data points.
UNDEFINED DIVIDE BY ZERO
Denominator cannot be zero in equa-
tion.
Make denominator a value other than
zero.
WINDOW TOO SMALL
Attempted to set time domain range
smaller than allowed
Re-enter larger time range.
OUT OF WINDOW RANGE
Attempted to set time domain range
larger than allowed
Re-enter values within allowed range.
Table B-1. Operational Error Messages (2 of 2)