Dsc74c, Dmcoil, Dss70a sampling system – Vaisala DM70 User Manual
Page 90

USER'S GUIDE____________________________________________________________________
Material Stainless
steel
AISI316
Weight 390
g
DSC74C
DSC74C is a sampling cell with DSC74B with DMCOIL
cooling/venting coil.
DMCOIL
Coil pipe size
6 mm
Connection to sampling cell With ISO G 1/4" and G 3/8"
thread adapter
Weight 130
g
DSS70A Sampling System
Measuring range
-60 ˚C T
d
... T
amb
(max T
amb
+40 ˚C T
d
)
Operating gases
Air, N
2
, non-toxic gases,
non-flammable, inert gases
Inlet/Outlet connection
1/4" SWAGELOK
Operating temperature
Ambient temperature
0 ... +40 ˚C (+32 ... +104 ˚F)
Gas temperature
max 200 ˚C (392 ˚F) with
2.0 m FEP tube at 20 ˚C
(68 ˚F)
Operating pressure
with pump
0.6 ... 1.2 bar
a
(8.7 ... 17.4 psi
a
)
pump disconnected
0 ... 20 bar
a
(0 ... 290 psi
a
)
Battery operation time for
pump
8 h continuous
Filter
7 mm inline filter, 1/4"
SWAGELOK SS-4F-7
Wetted parts
Stainless steel
Carrying case
ABC plastic
Weight
5.5 kg (12 lbs.)
88 __________________________________________________________________ M010091EN-F