Adjusting temperature, Temperature adjustment of dm70, One-point adjustment – Vaisala DM70 User Manual
Page 76
![background image](/manuals/408486/76/background.png)
USER'S GUIDE____________________________________________________________________
Adjusting Temperature
Temperature Adjustment of DM70
The temperature adjustment can be done using the one-point or two-
point adjustment. The one-point adjustment should be done within the
temperature range in which the device is most often used.
To go to the adjustment mode, do the following:
1. Press
the
Calibration button of the probe handle (see number 4
in Figure 1 on page 11) using a tool with a thin, sharp point, for
example, a small screwdriver. The indicator will switch into the
adjustment mode. To confirm calibration, press
OK.
2. Select
T
d/f
and press
Select.
3. Select
T and press
Select.
4.
To select the adjustment method, first press
Adjust and then
the adjustment method, either 1-point adjustment or 2-point
adjustment.
5.
Depending on your selection, continue from section One-Point
Adjustment below or Two-Point Adjustment on page 75.
One-Point Adjustment
1. After
selecting
1-point adjustment, press
Select to
continue.
2.
Set the probe to the reference temperature. You can follow the
stabilization from the graph display by pressing
Graph.
When the reading has stabilized, press
Ready.
3.
Use the arrow buttons to give the reference temperature value
and then press
OK.
4.
Confirm the adjustment by pressing
Yes. (If you press
No, the adjustment mode display returns and no changes are
made.) The adjustment has now been completed.
5.
To exit the adjustment mode, press
Back.
6.
To return to the basic display, press
Exit.
74 __________________________________________________________________ M010091EN-F