Absolute humidity (dmp74a probe recommended), Mixing ratio (dmp74a probe recommended), Relative humidity (dmp74a) – Vaisala DM70 User Manual
Page 86: Measurement environment
USER'S GUIDE____________________________________________________________________
Absolute Humidity (DMP74A Probe
Recommended)
Measurement range (typical) 0.5 ... 100 g/m
3
(0.2 ... 40 gr/ft
3
)
Accuracy
± (0.2 g/m
3
+ 10 % of
reading)
± (0.1 gr/ft
3
+ 10 % of
reading)
Mixing Ratio (DMP74A Probe Recommended)
Measurement range (typical) 0.2 ... 100 g/kg
(2 ... 700 gr/lbs)
Accuracy
± (0.1 g/kg + 12 % of
reading)
± (0.8 gr/lbs + 12 % of
reading)
Relative Humidity (DMP74A)
Measurement range
0 ... 100 %
Accuracy at +20 ˚C
RH < 5 %
± (0.025 % RH + 17.5 % of
reading)
RH > 5 %
± (0.8 % RH + 2 % of
reading)
Measurement Environment
Temperature
-10 ... +60 ˚C (+14 ... +140
˚F)
Pressure
DMP74A
0 ... 20 bar
a
(0 ... 290 psi
a
)
DMP74B
0 ... 20 bar
a
(0 ... 290 psi
a
)
DMP74C
0 ... 10 bar
a
(0 ... 150 psi
a
)
Sample flow rate
No effect on the dewpoint
accuracy
84 __________________________________________________________________ M010091EN-F