Setting Alarm Mode

The Sitter Elite has four (4) alarm modes. These allow you to select an alarm signal best suited

to patient and facility needs. The LED blinks green above the mode that the alarm is set to

when the alarm is on (fig. 6, page 10). The LED above the MUTE indicator blinks yellow when

the alarm is on and in “Mute” mode (fig. 6, page 10).

Mode Features:

When using a nurse call cable, ensure the nurse call cable is plugged in to

both the alarm and the wall jack before leaving the patient unattended. Verify that an

alert is received at the nursing station if the cable is unplugged from the wall jack.
NOTE: There will be no alert at the nursing station or at the bedside if the nurse call cable

is unplugged from the alarm.

To Change/Select Mode:

NOTE: “Voice Only” and “Mute” can be set only

when the nurse call cable is plugged into your alarm.

With this feature, all mode options are available.

Be sure to plug the nurse call cable into wall and

test. You MUST plug the nurse call cable into the

alarm AND the wall jack in order to activate nursing

station alerts.
Follow these steps to change or select alarm mode:

1. Lift the programming control protective cover

(fig. 4).

2. Ensure alarm is on and either sensor pad, chair

belt sensor, or magnet is attached to alarm.

Ensure pressure is on sensor pad, chair belt sensor

is connected, or magnet is on face plate.

3. Press MODE button on left side of alarm to scroll

through the mode selections (fig. 5).

TONE

Fig. 4

TONE

0
1
2

Fig. 5

VOICE & TONE: Depending on whether a custom voice recording is made, either the factory

default voice or a custom voice recorded message plays once, followed by

the selected alarm tone. The tone continues to play until alarm is placed on

hold or patient is repositioned with the magnet or sensor.

TONE:

This is the factory default mode. Selected alarm tone plays until alarm is

placed on hold or patient is repositioned with the magnet or sensor.

VOICE ONLY:

(This mode requires the nurse call cable to be plugged into your alarm and

the appropriate wall jack from your nurse call system). Custom voice or

factory default voice message plays repeatedly until alarm is placed on

hold or patient is repositioned with the magnet or sensor. This mode also

activates the facility nurse call system. Automatically switches to VOICE &

TONE if a nurse call cable is removed from the alarm (see page 18).

MUTE:

(This mode requires the nurse call cable to be plugged into your alarm

and the appropriate wall jack from your nurse call system). There is no

audible voice message or alarm tone at the patient’s bedside. This mode

activates the facility nurse call system. Automatically switches to VOICE &

TONE if the nurse call cable is removed from the alarm (see page 18).