Newlife, Intensity, Iv. 11. troubleshooting – AirSep MN139-1 User Manual

Page 69: Observations probable causes solutions, Airsep, Corporation

background image

NewLife

®

Intensity

F

AirSep

®

Corporation

IV. 11. troubleshooting

9-Gb

The 0-I button is in the ON

Power cable not plugged in

Check the cable connection.

position. The device does not

correctly. Mains power failure.

Reset the circuit breaker (2) if

operate. The continuous alarm

necessary by pressing. Check

sounds.

the fuses or circuit breaker

fitted on the premises.

Oxygen concentration indicator

Oxygen concentration is too

Contact your distributor.

remains lit yellow.

low.

The alarm test does not work.

Internal electrical fault.

Contact your distributor.

The 0-I button is ON and the

Pneumatic connection broken

Stop the device by pressing the

compressor is operating but

or compressor problem.

0-I button and contact your

there is no flow. The audible

distributor.

alarm sounds.

The 0-I button is ON, the

Internal electrical fault.

Stop the device and contact

compressor is operating, there

Pneumatic circuit fault.

your distributor.

is a flow but the audible alarm

sounds.

The compressor stops in mid-

Compressor thermal safety

Stop the device and wait for it

cycle, then starts again after a

device has been tripped. Fan

to cool down. Check that the

few minutes.

not working.

patient circuit is not obstructed.

Start up again. If the device

does not start, contact your

distributor.

The oxygen-enriched air flow is

Tube disconnected or humidifier

Check the gas administration

interrupted at the nasal

not tight.

circuit.

cannula outlet.

The flow at the nasal cannula

Pneumatic circuit problem.

Contact your distributor.

outlet is irregular.

observations

Probable causes

solutions

GB