Newlife, Intensity, Iv. 11. troubleshooting – AirSep MN139-1 User Manual
Page 69: Observations probable causes solutions, Airsep, Corporation

NewLife
®
Intensity
F
AirSep
®
Corporation
IV. 11. troubleshooting
9-Gb
The 0-I button is in the ON
Power cable not plugged in
Check the cable connection.
position. The device does not
correctly. Mains power failure.
Reset the circuit breaker (2) if
operate. The continuous alarm
necessary by pressing. Check
sounds.
the fuses or circuit breaker
fitted on the premises.
Oxygen concentration indicator
Oxygen concentration is too
Contact your distributor.
remains lit yellow.
low.
The alarm test does not work.
Internal electrical fault.
Contact your distributor.
The 0-I button is ON and the
Pneumatic connection broken
Stop the device by pressing the
compressor is operating but
or compressor problem.
0-I button and contact your
there is no flow. The audible
distributor.
alarm sounds.
The 0-I button is ON, the
Internal electrical fault.
Stop the device and contact
compressor is operating, there
Pneumatic circuit fault.
your distributor.
is a flow but the audible alarm
sounds.
The compressor stops in mid-
Compressor thermal safety
Stop the device and wait for it
cycle, then starts again after a
device has been tripped. Fan
to cool down. Check that the
few minutes.
not working.
patient circuit is not obstructed.
Start up again. If the device
does not start, contact your
distributor.
The oxygen-enriched air flow is
Tube disconnected or humidifier
Check the gas administration
interrupted at the nasal
not tight.
circuit.
cannula outlet.
The flow at the nasal cannula
Pneumatic circuit problem.
Contact your distributor.
outlet is irregular.
observations
Probable causes
solutions
GB