IV. useFuL INForMatIoN

IV. 1. accessories and spare parts

The accessories used with the NewLife

®

Intensity must:

- be oxygen compatible,

- be biocompatible,

- comply with the general requirements of the 93/42/EEC European Directive.

The connectors, tubes, nasal cannulas, probes or masks must be designed for oxygen therapy.

Contact your distributor to obtain these accessories.

remarks:

• The use of certain administration accessories which are not specified for use with this

concentrator may reduce its performance and void the manufacturer’s responsibility.

IV. 2. Materials in direct or indirect contact with the patient

Concentrator casing ............................. Valtra/ABS/Polystyrene

Mains cable ................................................. PVC

Dust filter ..................................................... Polyester

ON/OFF switch ............................................ Thermoplastic

Casters ........................................................ Nylon

Flow adjustment knob .................................. ABS/Polycarbonate

Gas outlet .................................................... Chrome Plated Brass

Printed labels ............................................... Lexan

IV. 3. operating principle

The compressor sends filtered ambient air to a group of valves, which allows compressed air to

pass to the column in production. The columns contain a molecular sieve, whose function is to

adsorb the nitrogen and thus allow oxygen to pass. The oxygen-enriched air is then directed to a

pressure reducing valve through the flow control valve to the oxygen outlet fitting.

During this time, the column which is being “regenerated” is connected to the ambient air and a

current of oxygen enriched air is passed through it (from the column “in production”). In this way,

when one column is in production, the other is in a nitrogen desorption or “regeneration” phase. The

oxygen enriched air finally passes through a product filter situated prior to the oxygen therapy outlet.