Medical equipment needs special precautions regarding EMC and need to be installed and put into service

according to the EMC information provided in this section.

Guidance and manufacturer’s declaration ± electromagnetic immunity

The FreeStyle is intended for use in the electromagnetic environment specified below. The customer or the user of the FreeStyle should assure
that it is used in such an environment.

I

MMUNITY

test

IEC 60601

test level

Compliance level

Electromagnetic environment ±

guidance

Electrostatic
discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30 %.

Electrical fast
transient/burst

IEC 61000-4-4

± 2 kV for power
supply lines

± 1 kV for input/output
lines

± 2 kV for power
supply lines

Not Applicable

Mains power quality should be that of a typical commercial or
hospital environment.

Surge

IEC 61000-4-5

± 1 kV line to line

± 2 kV line to earth

± 1 kV line to line

± 2 kV line to earth

Mains power quality should be that of a typical commercial or
hospital environment.

Voltage dips, short
interruptions and
voltage variations
on power.

IEC 61000-4-11

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(>30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5s

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 s

Mains power quality should be that of a typical commercial or
hospital environment. If the user of the FreeStyle requires
continued operation during power mains interruptions, it is
recommended that the FreeStyle be powered from an
uninterruptible power supply (UPS) or a battery.

Power frequency
magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.

NOTE U

T

is the a.c. mains voltage prior to application of the test level.