AirSep offers the OxySafe as an optional accessory. This is intended to be used in conjunction with the FreeStyle
concentrator. For customers in regions requiring compliance to EN ISO 8359:1996-Ammendment1:2012, this accessory will
meet this need.

The OxySafe is a thermal fuse to stop the flow of gas in the event that the downstream cannula or oxygen tubing is ignited and
burns to the OxySafe. It is placed in-line with the nasal cannula or oxygen tubing between the patient and the oxygen outlet of
the Visionaire.

For proper use of the OxySafe, always refer to the manufacturer’s instructions (included with each OxySafe kit).

AirSep offers an OxySafe kit that includes OxySafe with 2” of tubing to connect to the oxygen outlet; PN 20629671

Important Safety Rules for Optional AirBelt

 

 

 

For oxygen concentrators equipped with the optional Airbelt :

1. The incorrect use of AirBelt can cause the battery to get hot, ignite, and can cause serious injury. Be sure

not to pierce, strike, step on, or drop the battery, or otherwise subject the battery to strong impacts or
shocks.

2. Replace safety cap on AirBelt cord when not in use.

 

 

For oxygen concentrators equipped with batteries:

1. When connected to AC or DC power, the unit’s battery charges until it reaches full capacity, either while

the unit is operating or turned off.

2. If the internal battery fully depletes and the Oxygen Concentrator shuts down, the unit cannot be restarted

with the optional AirBelt. Should this occur, connect your Oxygen Concentrator to its AC or DC power
supply for a short period of time in order to provide sufficient internal battery power to start the unit.
AirBelt can then be connected to provide additional run time.

 

 

For oxygen concentrators equipped with the optional Airbelt:

1. The AirBelt does not need to be fully discharged before recharging. It is recommended to charge it after

each use.

2. Depending upon the temperature of the AirBelt, it can take several minutes for the charging cycle to start

after connecting to power. This is a normal condition and is intended for safe charging.