AirSep MN123-1 F User Manual
Page 32
MN123-1 rev F
28
How to Respond to FreeStyle’s Alarm/Light Indicators
Status
Audible Alarm
Light
Indicates
Action
Indicator
Brief, continuous at
start–up
(Green) pulse;
continuous light
FreeStyle has been
turned on
You may begin to operate your
FreeStyle unit.
Indicator
No
(Green) pulse;
intermittent light
FreeStyle is delivering
oxygen as a pulse flow.
Continue using FreeStyle normally.
Indicator
No
Level indicator
Battery charge level.
Charge as indicated.
Indicator
No
(Yellow) continuous
light
Service required.
Return unit to Equipment Provider
for inspection/service.
Alarm
Continuous:
Beep
(Red) alarm;
continuous light
No breath detected by the
unit for a predetermined
time period.
Check the cannula connection. Ensure
that you are breathing through your
nose. If the alarm persists, contact
your Equipment provider.
Alarm
Intermittent:
Beep, beep, beep.
25% (Yellow)
BATT; battery indicator.
Battery voltage is too low
to operate FreeStyle.
Connect the FreeStyle unit into a DC
outlet or an AC outlet immediately.
Alarm
Rapid intermittent:
Beep, beep, beep
(Red) alarm;
intermittent light
Breathing rate is
exceeding the capacity of
the FreeStyle unit.
Reduce activity, then if necessary use
another source of oxygen as available.
Contact your Equipment Provider.
Alarm
Rapid intermittent:
Beep, beep, beep
(Red) alarm;
continuous light
General malfunction of the
FreeStyle unit has
occurred.
Turn off the unit. Change to another
source of oxygen, and contact your