beautypg.com

Freestyle specifications – AirSep MN123-1 F User Manual

Page 40

background image

MN123-1 rev F

36

FreeStyle Specifications

Oxygen Concentration:*

1-3 pulse settings; equivalent to a continuous flow of 90% oxygen +5.5/-3%

Flow Rate Settings: 1 ,2 & 3 LPM (equivalent to continuous flow)
Settings 1 through 3 ±10%

Dimensions:

8.6 in. high x 6.1 in. wide x 3.6 in. deep (21.8 cm high x 15.5 cm wide x 9.1 cm
deep)

Weight:

4.4 lb (2.0 kg); 1.8 lb (.8 kg) optional AirBelt


Power:

Universal AC/DC Power Supply (100-240VAC 1.75A Max 50/60Hz)
(11-16VDC 8.0A Max)

Battery duration
(Rechargeable lithium
Battery)

3 LPM – 2 hours; 2 LPM – 2.5 hours; 1 LPM – 3.5 hours
Optional AirBelt when combined with the internal battery:
3 LPM – 5 hours; 2 LPM – 6 hours; 1 LPM – 10 hours

Battery recharge time:

3.5 hours; optional AirBelt: 3 hours

Warm-up time:

2 minutes

Battery cycle life:

Approximately 300 cycles, then 80% capacity or below.

Audible alarms and pulse
visual indicators:

Start-up – audible and visual (GREEN indicator light);
Pulse flow – visual (GREEN light);
Battery condition – battery level (GREEN indicator lights);
Cannula disconnect- audible and visual (RED alarm light);
System overdraw – audible and visual (RED warning and alarm lights);
High and low pressure – audible and visual (RED alarm light);
Service required: visual (YELLOW alarm light)

Sound:

38 dBA at 1 setting: 41 dBA at 2 setting: 44 dBA at 3 setting

**Temperature range:

Operational temperature: 41°F to 104°F (5°C to 40°C)

(Up to 95% RH (non-condensing)

Storage temperature:

-4°F to 140°F (-20°C to 60°C)

**Max Operational
Altitude:

up to 12,000 ft (3,657.6 m) (483mmHg) Higher altitudes may affect performance

* Based on an atmospheric pressure of 14.7 psi (101 kPa) at 70°F (21°C)
**Operating outside of these operational specifications can limit the concentrator's ability to meet Oxygen Concentration specification at higher liter flow rates.