Using more than one controller, Starting run-time mode, Run-time measurement sequence – Agilent Technologies Multiple Channel Controller 85330A User Manual
Page 88
6-14
85330A Multiple Channel Controller
Programming
To Use Run-Time Control Mode
Using More than One
Controller
More than one multiple channel controller may be used in a system.
Multiple controllers are usually used when the distance to a switch control
unit exceeds the maximum length allowed. The REMOTE 1 and REMOTE 2
connectors on the rear panel allow connections to multiple controllers. See
for more information on these connectors.
To set up and control the remote controller, use these commands:
•
RUNTime:CONTroller
•
RUNTime:TIMEout:REMote
•
RUNTime:SWITch:TRIGger
The remote controllers can be triggered using the commands:
•
RUNTime:SOURce:SOURCE1:TRIGger
•
RUNTime:SOURce:SOURCE2:TRIGger
Starting Run-Time
Mode
Issue the
RUNT:INIT:IMM
command to initiate the run-time operation. This
command passes the hardware control to the 85330A and allows it to accept
and issue triggers.
OUTPUT 70918;"RUNT:INIT:IMM;”
Once the sequence is initiated the 85330A either:
•
Waits for an event trigger before issuing a trigger to the receiver,
OR:
•
Immediately issues a trigger to the receiver.
This is set using the
RUNT:EVEN:TRIG
command.
Run-Time
Measurement
Sequence
Refer to
Figure 6-2, “Run-Time Flow Chart with Programming Command
. The chart shows the sequence of a run-time
measurement. Programming commands applicable to each function are
shown. All programming commands must be executed before run-time is
initiated. Here is a description of the run-time measurement sequence:
Idle
This is the idle state of the 85330A. This also the state the unit enters if an
error occurs during run-time. Run-time mode is started when
RUNT:INIT:IMM
is
executed.
Select Switch State
The first switch state is selected for Port 1 or Port 2, as defined with the
RUNT:SWIT:SCAN
command. A delay now occurs to allow the switch to settle,
defined with
RUNT:SWIT:DEL
.