Aux 1 and aux 2 output voltage test, When finished with all tests – Agilent Technologies Multiple Channel Controller 85330A User Manual
Page 48
![background image](/manuals/39387/48/background.png)
2-18
85330A Multiple Channel Controller
Performance Verification
AUX 1 and AUX 2
AUX 1 and AUX 2
Output Voltage Test
The following test will determine if the rear panel ports successfully output
the correct DC switch drive levels.
1. On the oscilloscope press:
[RECALL] [1]
[TRIG] {AUTO}
[CLEAR DISPLAY]
2. Press the following on the multiple channel controller:
[Clear Instr] [Select Instr] {85330A}
3. Connect a BNC cable between the AUX 1 connector and channel 1 of
the oscilloscope.
a.
Press:
{CLOSE} {AUX1} {0}
or
[f1]
The oscilloscope should show a trace of about 0 Vdc.
b.
Press:
{1}
or
[f2]
The oscilloscope should show a trace of about 4 Vdc.
4. Connect a BNC cable between the AUX 2 connector and channel 1 of
the oscilloscope.
a.
Press:
[Clear Instr] {CLOSE} {AUX2} {0}
or
[f1]
The oscilloscope should show a trace of about 0 Vdc.
b.
Press:
{1}
or
[f2]
The oscilloscope should show a trace of about 4 Vdc.
When Finished with
All Tests
All of the performance verification tests are now complete. Reconnect the
85330A system in the original measurement system configuration. Refer to