7-4
85330A Multiple Channel Controller
Remote Programming Command Reference
Common Commands
*OPC?
The operation complete command. This can be used to measure timing of a
run-time sequence. The command
*OPC?
returns a value of 1 when evoked.
Prior commands are executed before the completion of
*OPC?
.
Syntax
*OPC?
Example
10 OUTPUT 70918;”RUNT:EVEN:COUN 1;”
20 OUTPUT 70918;”TRIG:SOUR IMM;”
30 TIME_START=TIMEDATE
40 OUTPUT 70918;”INIT:IMM;*OPC?”
50 ENTER 70918;OPC_NUM
60 TIME_STOP=TIMEDATE
70 PRINT “MEASUREMENT TIME (SEC)=”;TIME_STOP-TIME_START
*RST
A soft reset of the E1330A/B and 85330-60002 card. It is executed at
power-up. This command does not clear the error stack.
Default
All card address are set to channel 1.
Syntax
*RST
Example
OUTPUT 70918;”*RST”
NOTE
A
*RST
will produce a trigger signal on the SRC 1 TRIG, SRC2 TRIG, and
RCVR TRIG lines.