Troubleshooting the switch control unit, Troubleshooting the rf switch, Troubleshooting the cables – Agilent Technologies Multiple Channel Controller 85330A User Manual
Page 140
85330A Multiple Channel Controller
9-11
Service
Troubleshooting
Troubleshooting the
Switch Control Unit
Run the test
“Verifying the Switch Control Unit” on page 2-12
NOTE
Check to make sure that the serial number prefixes of the Switch Control
Unit and RF switch match. They are produced at the factory this way. It is
acceptable to use switches and SCUs with different serial numbers. But if
there is a problem, it may be because a newer switch is used with an older
SCU, or an older switch is used with a newer SCU.
Troubleshooting the
RF Switch
Run the test
“Verifying the RF Switches” on page 2-13.
Troubleshooting the
Cables
The cables will be checked during “Verifying the Switch Control Unit” and
“Verifying the RF Switches” tests in
Chapter 2, “Performance Verification.”
If a cable is suspected, use a voltmeter to check continuity between the
connectors on each end of the cable. See
wiring information for each set of cables.
Table 9-3
SWITCH PORT 1/2 Connector Pinout
D-Type
Connector Pins
Function
D-Type
Connector Pins
Function
1
Bit 6 HI
14
Bit 2 HI
2
Bit 6 LO
15
Bit 2 LO
3
Bit 7 HI
16
Bit 3 HI
4
Bit 7 LO
17
Bit 3 LO
5
+
12 Vdc
18
Bit 4 HI
6
+
12 Vdc
19
Bit 4 LO
7
GND
20
Bit 5 HI
8
GND
21
Bit 5 LO
9
GND
22
+
12 Vdc
0
Bit 0 HI
23
−
12 Vdc
11
Bit 0 LO
24
GND
12
Bit 1 HI
25
GND
13
Bit 1 LO