Features and Functions
2
U2761A User’s Guide
79
ASK Rate
The ASK rate is the rate at which the output amplitude
“shifts” between two preset amplitudes. The range of the
ASK rate is 2 mHz to 100 kHz. The default ASK rate value is
10 Hz.
Soft Front Panel Operation
On the ASK Rate panel as shown in
, input the
desired ASK rate value and select the unit from the drop
down list.
Remote Interface Operation
ASKey:INTernal:RATE
Generate ASK Modulation Using SCPI Commands
Example 1
-> *CLS
; *RST
//To reset the U2761A to
default power
-on state, this
command can be ignored if
this operation is not
required.
-> ASK:STAT ON
//Enables ASK modulation.
-> FUNC SIN
//Sets the carrier waveform to
Sine wave.
-> FREQ 1000
//Sets the carrier frequency to
1 kHz.
-> VOLT 5 VPP
//Sets the output amplitude to
5 Vpp.
-> VOLT:OFFS 0
//Sets the output offset to 0.
-> ASK:INT:RATE 10
//Sets the ASK rate to
10 Hz.
-> OUTP ON
//Turns on output.