Output units – Agilent Technologies AGILENT U2761A User Manual
Page 53

Features and Functions
2
U2761A User’s Guide
35
Output Units
This configuration applies to output amplitude only. At
power
-on, the units for output amplitude are volts
peak
-to-peak.
• The output units consist of Vpp, Vrms, or dBm. The
default unit is Vpp.
• The unit setting is stored in volatile memory. The units
are set to “Vpp” upon power
-off or after a remote
interface reset.
• The output units for amplitude cannot be set to dBm if
the output termination is currently set to
“high impedance”. The units are automatically converted
to Vpp.
Soft Front Panel Operation
On the Amplitude panel as shown in
, input the
desired amplitude value and select the unit from the drop
down list.
Remote Interface Operation
VOLTage:UNIT {VPP|VRMS|DBM}
See also other documents in the category Agilent Technologies Hardware:
- HP 16500B (38 pages)
- E4400-60230 (18 pages)
- Option H48 Multiport Test Set Z5623A (138 pages)
- hard disk drive kit e5052-90202 (13 pages)
- FIXTURE U7242A (10 pages)
- FS2010 (36 pages)
- USB/GPIB Interface Converter 82357A (70 pages)
- 34399A RS-232 (3 pages)
- 1100 Series (28 pages)
- N9340A (161 pages)
- HSSI WanProbe J3916A (179 pages)
- 37907A (16 pages)
- N1022A (24 pages)
- Agilent E6474A (184 pages)
- 82357B (34 pages)
- Agilent E1345A (5 pages)
- 85225F (148 pages)
- 8510C (16 pages)
- Wireless Network Optimization Platform E6474A (194 pages)
- J3919A (189 pages)
- 85032B/E 50 (55 pages)
- 8510 (32 pages)
- J6800A (16 pages)
- E6474A (14 pages)
- N2610A (248 pages)
- N6314A (7 pages)
- 85107B (20 pages)
- 8753ET (28 pages)
- N9360A (44 pages)
- Agilent 82350B (51 pages)
- Agilent 4396B (124 pages)
- LXI (8 pages)
- E1439 (246 pages)
- U2802A (67 pages)
- Agilent Hard Disk Drive Kit E5071BU (12 pages)
- U2781A (47 pages)
- E1301B (244 pages)
- Multiple Channel Controller 85330A (166 pages)
- E6701A (32 pages)
- N2620A (279 pages)
- E5810A (159 pages)
- E5805A (47 pages)