beautypg.com

Carrier frequency, Table – Agilent Technologies AGILENT U2761A User Manual

Page 96

background image

78

U2761A User’s Guide

2

Features and Functions

Remote Interface Operation

FUNCtion {SINusoid|SQUare|RAMP|USER}

You can also use the APPLy command to select the function,
frequency, amplitude, and offset.

Carrier Frequency

The maximum carrier frequency depends on the function
selected as shown below. The default is 1 kHz for all functions.

Only Square wave with a 50% duty cycle is available for the
Internal ASK modulating waveform.

Soft Front Panel Operation

On the Frequency panel as shown in

Figure 2- 3

, input the

desired frequency value and select the unit from the drop
down list.

Remote Interface Operation

FREQuency

You can also use the APPLy command to select the function,
frequency, amplitude, and offset.

Table 2-9 Carrier frequency for ASK

Function

Minimum frequency

Maximum frequency

Sine

1

μHz

20 MHz

Square

1

μHz

20 MHz

Ramp, Triangle

1

μHz

200 kHz

Arbitrary

1

μHz

200 kHz