External triggering – Agilent Technologies AGILENT U2761A User Manual
Page 105
Features and Functions
2
U2761A User’s Guide
87
Remote Interface Operation
TRIGger:SOURce BUS
When the BUS source is selected, send the TRIG or *TRG
command to trigger the U2761A.
External Triggering
In the External trigger mode, the U2761A accepts a
hardware trigger applied to the device Trig In connector. The
U2761A initiates one sweep each time Trig In receives a TTL
pulse with the specified edge.
Also refer to
Soft Front Panel Operation
On the Trigger Setup panel as shown in
, select
External from the drop down list
at the Source
panel.
To specify whether the U2761A is triggered on the rising or
falling edge, select
Positive
for triggering on the
rising edge and
Negative
for triggering on the falling
edge.
Remote Interface Operation
TRIGger:SOURce EXTernal
Use the following command to specify whether the U2761A
is triggered on the rising or falling edge.
TRIGger:SLOPe {POSitive|NEGative}