3.0 Electrical Installation

The pin out for the D-sub, 15 pin connector is shown in Figures 5, page 6.
A schematic diagram of the recommended electrical connections to the
D-sub, 15 pin connector is shown in Figure 6, page 6.

The pin out for the 6-Position Screw Terminal Connector is shown in
Figure 7, page 7. A schematic diagram of the recommended electrical
connection to the 6-position Screw Terminal Connection is shown in
Figure 8.

The model 760 operates from a

±

15 VDC regulated (

±

5%) power supply.

The model 760 has either a 0-5 VDC output or 0-10 VDC output.

Note 1: The ground of any external power supply and readout system
should be the same as the transducer ground (chassis ground) to minimize
any possible ground loops which may effect the performance and stability
of the transducer.

Note 2: The Model 760 meets CE mark requirements and complies with
EMC Directive 89/336/EEC. To ensure compliance when installed, an overall
metal braided shielded cable (and metal shielded connectors for the D sub
connector option) connected to chassis ground at both ends is required.

5