2.0 Mechanical Installation
Remove all packaging material and the protective flange cover and visually
check the Model 760. If the Model 760 appears damaged, notify Setra
Systems or your supplier immediately. Retain packaging materials for
inspection. Do not use if damaged. If the Model 760 is not going to be used
immediately, then replace the protective flange cover and store in an area
where the temperature range is controlled between -50 to +125
°
C.
The Model 760 can be mounted in any orientation on the vacuum system.
To avoid the buildup of debris or condensable material in the measure-
ment cavity of the Model 760 (which may cause measurement errors), we
recommend that you install the Model 760 vertically with the tube facing
down. Outline drawings showing the external dimensions are shown in
Figures 1 - 4.
To connect the Model 760 to your system use the appropriate hardware for
the type of fitting:
• Use a Cajon® Ultra Torr type of compression coupling to connect to the
O.5" OD tube.
• Use an O-ring/centering ring and clamp to connect to the NW16, 25 and
40 flange options.
Note: a stepped O-ring carrier may be used to connect the NW16 flange to an NW10 flange
on the system.
• Use a Male 8 VCR™ Style Face Seal Fitting and sealing washer to connect
to the female swivel 8 VCR™ Style Face Seal Fitting.
• Use a Male 0.5" Swagelock™ fitting/union to connect to the 0.5"
Swagelock™ Nut fitting.
• Use an O-ring and clamp to connect to the 1.5" Triclover™ style sanitary
flange.
Note: Tighten threaded fittings in accordance with the manufacturer’s