7.0 Specifications

Performance Data:
Accuracy (RSS)

1

:

<

±

0.25% of Reading

Optional <

±

0.15% of Reading

Resolution:

0.01% FS

Thermal effects

Compensated Range:

0 to +50

°

C

Zero Shift:

<

±

0.005% FS /

°

C

Span Shift:

<

±

0.027% FS /

°

C

Proof Pressure:

45 psia

Operating Temperature:

0 to +50

°

C

Storage Temperature:

-50 to +125

°

C

Electrical Data:
Connector:

15 pin D sub or 6 position terminal strip

Excitation:

±

15 V DC, regulated

±

5%

Output:

0 - 10 V DC or 0 - 5 V DC

Output Maximum Load:

<10 K

load

Power Consumption:

< 0.5 Watts (<15 mA)

Time Constant:

< 20 ms

EMC Performance:

Complies with EMC Directive 89/336/EEC

Physical Description:
Case:

Aluminum Alloy

Vacuum Fittings:

0.5" OD Tube; other fittings available; see
reorder data page

Wetted Materials

2

:

Inconel®

Measurement cavity volume

3

: <6 cm

3

Weight:

260 g

Notes:

1. Root sum of the squares (RSS) of linearity, hysteresis and non-repeatability. Accuracy is
expressed as % of reading. However, near Zero, the accuracy is limited by the resolution of the
instrument. So, the accuracy is more correctly stated as the greater of

±

0.25% reading or

±

0.01%

FS. (For the optional accuracy, this becomes the greater of

±

0.15% reading or

±

0.01% FS).

2. Wetted material is for 0.5" tube option only. Other flange options will add stainless steel.

3. Maximum cavity volume including the 0.5" OD tube volume of 0.26 in

3

(4.28 cm

3

).

11