beautypg.com

E-m-hf8-v2_10 – ROTRONIC HF8 User Manual

Page 33

background image

E-M-HF8-V2_10

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroFlex HF8 Humidity Temperature
Transmitter: User Guide

Instruction Manual

Document Type

Page 33 of 35

Document title

© 2012; Rotronic AG E-M-HF8-V2_10

Configurable analog outputs

HF8

Output 1

Can be made to correspond to any parameter and probe

Parameter

User configurable with the ROTRONIC HW4 software

Scale

User configurable with the ROTRONIC HW4 software

Output 2

Can be made to correspond to any parameter and probe

Parameter

User configurable with the ROTRONIC HW4 software

Scale

User configurable with the ROTRONIC HW4 software

Output 3

Can be made to correspond to any parameter and probe

Parameter

User configurable with the ROTRONIC HW4 software

Scale

User configurable with the ROTRONIC HW4 software

Output 4

Can be made to correspond to any parameter and probe

Parameter

User configurable with the ROTRONIC HW4 software

Scale

User configurable with the ROTRONIC HW4 software

Output 1, 2, 3 and 4

Signal type

0…20 mA
4… 20 mA
0… 1 V
0… 5 V
0… 10 V
(user configurable with the ROTRONIC HW4 software)

User configurable scaling limits

-999.99 … +9999.99 engineering units

Short circuit tolerant

Yes

Maximum external load

500 Ω (current output)

Minimum external load

1000 Ω (voltage output)

0 Ω (current output)

Configurable relay outputs

HF8

Relay 1

Can be made to correspond to any parameter and probe

Parameter

User configurable with the ROTRONIC HW4 software

Operation

User configurable with the ROTRONIC HW4 software

Relay 2

Can be made to correspond to any parameter and probe

Parameter

User configurable with the ROTRONIC HW4 software

Operation

User configurable with the ROTRONIC HW4 software

Relay 3

Can be made to correspond to any parameter and probe

Parameter

User configurable with the ROTRONIC HW4 software

Operation

User configurable with the ROTRONIC HW4 software

Relay 4

Can be made to correspond to any parameter and probe

Parameter

User configurable with the ROTRONIC HW4 software

Operation

User configurable with the ROTRONIC HW4 software

Relays 1 , 2, 3 and 4

Relay type

SPDT (single pole double throw)

Maximum switching voltage

250 VAC 1)

Maximum switching current

6 A (AC) 1)

1) Resistive load