4 frequently used settings, 1 unit system, 2 date and time – ROTRONIC HF8 User Manual
Page 25: Frequently used settings, E-m-hf8-v2_10, 4 frequently used settings 7.4.1
E-M-HF8-V2_10
Rotronic AG
Bassersdorf, Switzerland
Document code
Unit
HygroFlex HF8 Humidity Temperature
Transmitter: User Guide
Instruction Manual
Document Type
Page 25 of 35
Document title
© 2012; Rotronic AG E-M-HF8-V2_10
Main Menu
Menu Items
Submenu
Items
Selections or
Information
Notes
Data Logging
o
Recording
o
Samples
o
Interval
o
Mode
o
Recording (ON / OFF)
o
Nbr of records
o
Log Interval
o
Mode:
Start Stop / Loop
Start Recording
Stop Recording
Data logging by the HF8
(max. 10,000 H+T records
per probe)
Settings
Cannot be
changed while
the HF8 is
recording data
Interval
Log interval: 5 sec to 1 h
Mode
Start-Stop / Loop
Start- Stop: the recording
stops when the memory is
full
Loop: when the memory is
full the oldest record is
dumped to make room for
the new record
Probe 1
ON / OFF
Probe 2
ON / OFF
NOTE: use the MENU key to go back one step from any sub-menu or to exit the entire menu (this may require
several key presses).
7.4 Frequently used settings
7.4.1
Unit system
Press the MENU key and select Device Settings > Local Settings > Unit Sys. Press ENTER to activate the
Unit Sys menu item, use the [+] or [-] key to change the unit system. Press ENTER to confirm and press
MENU to exit.
The HW4 software can also be used to change the unit system.
7.4.2
Date and time
Press the MENU key and select Device Settings > Date or Time. Press ENTER to activate either the Date or
the Time menu item. Use the [+] or [-] key to change the Date or the Time. After each change, the cursor
moves to the right. When done, press ENTER to confirm and press MENU to exit.
To change either the date or the time format, Press the MENU key and select Device Settings > Local
Settings > Date Fmt or Time Fmt. Press ENTER to activate either the Date Fmt or the Time Fmt menu item,
use the [+] or [-] key to change the Date or the Time format. When done, press ENTER to confirm and press
MENU to exit.
The HW4 software can also be used to set the clock of the HF8 to the PC date and time.
7.4.3
Select the calculated parameter for a probe input
The calculated parameter is available only when the input is set for a digital HygroClip 2 probe. Press the
MENU key and select Device Settings > Input 1 or Input 2 > Calc. Press ENTER to activate the Calc sub-
menu, use the [+] or [-] key to select the calculated parameter. Press ENTER to confirm and press MENU to
exit.